Clinical evaluation of the pressure phosphene tonometer in patients with glaucoma
  1. E Rietveld,
  2. D A van den Bremer,
  3. H J Völker-Dieben
  1. Department of Ophthalmology, VU University Medical Center, Amsterdam, Netherlands
  1. Correspondence to: Eelco Rietveld MD, Department of Ophthalmology, VU University Medical Center, PO Box 7057, 1007 MB Amsterdam, Netherlands; e.rietveldvumc.nl

Abstract

Aim: To evaluate the reliability of the pressure phosphene tonometer in comparison with the Goldmann applanation tonometer.

Methods: 45 consecutive patients with glaucoma (78 eyes) participated in the study. Eyes with previous eye surgery, a documented peripheral visual field defect, a refractive error of more than 5 dioptres, and patients who were unable to understand the procedure were excluded from the study. Intraocular pressure was measured with a pressure phosphene tonometer by one examiner and with a Goldmann applanation tonometer by two other examiners no more than 15 minutes apart. A second series of measurements was performed several weeks later on 34 patients (59 eyes). There was no communication between examiners or between examiner and patient regarding test results.

Results: No statistically significant correlation was found between the applanation tonometry values and those obtained with a pressure phosphene tonometer.

Conclusion: The pressure phosphene tonometer is not suitable for reliably measuring intraocular pressure.

  • AT, applanation tonometer
  • PT, phosphene tonometer
  • glaucoma
  • phosphene tonometer
  • AT, applanation tonometer
  • PT, phosphene tonometer
  • glaucoma
  • phosphene tonometer