Low-cost, smartphone-based frequency doubling technology visual field testing using a head-mounted display

Abstract

Background Current visual field screening machines are bulky and expensive, limiting their accessibility, affordability and use. We report the design and evaluation of a novel, portable, cost-effective system for glaucoma screening using smartphone-based visual field screening using frequency doubling technology (FDT) and a head-mounted display.

Methods Nineteen eyes of 10 subjects with new-onset or chronic primary open angle glaucoma were tested and compared with the Humphrey Zeiss FDT and the newly developed Mobile Virtual Perimetry (MVP) FDT with the C-20 testing pattern. Mann-Whitney, Bland-Altman and linear regression analyses were performed to assess statistical difference, agreement and correlation, respectively, between the two devices.

Results The average age of the participants was 58±15 years. No statistically significant difference was found between the MVP FDT and the Humphrey Zeiss FDT (p>0.05). Bland-Altman and linear regression analyses demonstrated good agreement and correlation between the two devices.

Conclusion The MVP FDT is a low-cost, portable visual field screening device that produces comparable results to the Humphrey Zeiss FDT and may be used as an easily accessible screening tool for glaucoma.

  • Glaucoma
  • Field of vision
  • Telemedicine

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Linked Articles

  • At a glance
    Keith Barton James Chodosh Jost B Jonas