Microaneurysms visualisation using five different optical coherence tomography angiography devices compared to fluorescein angiography

Abstract

Background To compare fluorescein angiography (FA) and five different optical coherence tomography angiography (OCTA) devices and to test their reproducibility in the evaluation of retinal microaneurysms (MAs) secondary to diabetic retinopathy (DR).

Methods On the same day, patients with DR were imaged with FA and five OCTA devices: prototype Spectralis OCTA, prototype PlexElite, RTVue XR Avanti, AngioPlex and DRI OCT Triton. For all OCTA devices, a 3×3 volume scan pattern was performed. MAs were evaluated for the superficial capillary plexus (SCP) and deep capillary plexus (DCP).

Results Twenty eyes of 15 patients with DR were included. FA counted a significantly higher number of MAs compared to OCTA devices. Spectralis OCTA obtained a significantly higher number of MAs compared to PlexElite, RTVue XR Avanti, AngioPlex and DRI OCT Triton (p<0.0001). PlexElite and AngioPlex showed a greater number of MAs in the SCP, Spectralis OCTA, RTVue XR Avanti and DRI OCT Triton in the DCP. Higher sensitivity (43.3%) but lowest specificity (54.4%) was observed for Spectralis OCTA compared to other devices. The higher specificity (78.5%) and positive predictive value (83.3%) were observed for DRI OCT Triton.

Conclusions FA remains the best imaging modality to visualise retinal MAs. Spectralis OCTA was able to detect more MAs compared to other devices, likely due to the higher number of B-scans in the scanned area as well as due to the higher number of repeated B-scans. The high variability between OCTA devices should be taken into account for future clinical trials as in clinical practice.

  • Retina
  • Imaging
  • Inflammation
  • Immunology
  • Macula
  • Treatment Lasers
  • Diagnostic tests/Investigation
  • Treatment Medical
http://creativecommons.org/licenses/by-nc/4.0/

This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/.

View Full Text

Statistics from Altmetric.com

Footnotes

  • SP and FC contributed equally.

  • Contributors SP and FC have contributed to study conception, design, analysis, data acquisition, interpretation of data, draf and revision, and final approval. MC, DM and FZ have contributed to data acquisition, revision and final approval. GS have contributed to interpretation of data, draft and revision, and final approval.

  • Funding This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.

  • Competing interests None declared.

  • Data sharing statement All data relevant to the study are included in the article or uploaded as supplementary information.

  • Provenance and peer review Not commissioned; externally peer reviewed.

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.