RT Journal Article SR Electronic T1 Reproducibility and accuracy of measurements with a hand held autorefractor in children JF British Journal of Ophthalmology JO Br J Ophthalmol FD BMJ Publishing Group Ltd. SP 941 OP 948 DO 10.1136/bjo.81.11.941 VO 81 IS 11 A1 Erin M Harvey A1 Joseph M Miller A1 L Keith Wagner A1 Velma Dobson YR 1997 UL http://bjo.bmj.com/content/81/11/941.abstract AB AIM To determine reproducibility and accuracy of the Nikon Retinomax autorefractor when used with children who were made cycloplegic. METHODS Autorefraction and retinoscopy or subjectively refined retinoscopy (where, under the patient’s direction, the refraction was varied until the best visual acuity was achieved) were performed on the right eye of 47 children, age 11–93 months. Autorefraction was performed using the Nikon Retinomax, which provides up to eight measured values of refractive error and one representative measurement of refractive error. RESULTS Autorefractor measurements were successfully obtained from 7/9 children age 3 years or younger, and from all older children. Vector methods were used to calculate differences. Retinomax reproducibility averaged 0.43 D. Unbiased Retinomax and retinoscopy measurements differed by an average of 0.82 D. Unbiased Retinomax and subjectively refined retinoscopy differed by an average of 1.03 D. CONCLUSIONS Reproducibility of Retinomax measured values in children is comparable with reproducibility of retino scopy, subjective refraction, and autorefraction measurements in adults. Agreement between Retinomax and retinoscopy and agreement between Retinomax and subjective refinement in children is comparable with agreement between autorefraction and subjective refraction in adults. The study indicates that the Retinomax is a useful instrument for measuring refractive errors in young children.