RT Journal Article SR Electronic T1 Clinical evaluation of the MacuScope macular pigment densitometer JF British Journal of Ophthalmology JO Br J Ophthalmol FD BMJ Publishing Group Ltd. SP 328 OP 331 DO 10.1136/bjo.2009.167213 VO 94 IS 3 A1 Bartlett, Hannah A1 Acton, Jennifer A1 Eperjesi, Frank YR 2010 UL http://bjo.bmj.com/content/94/3/328.abstract AB Background/aims The MacuScope uses a psychophysical technique called heterochromic flicker photometry to measure macular pigment optical density (MPOD). Our aim was to determine the measurement variability (noise) of the MacuScope.Methods Thirty-eight normally sighted participants who ranged in age from 19 to 46 years (25.7±7.6 years) were recruited from staff and students of Aston University. Data were collected by two operators, HB and JA, in two sessions separated by 1 week in order to assess test repeatability and reproducibility.Results The overall mean MPOD for the cohort was 0.47±0.14. There was a significant negative correlation between MacuScope MPOD readings and age (r=−0.368, p=0.023). Coefficients were 0.45 and 0.58 for repeatability, and 0.49 and 0.36 for reproducibility. For each pair of results, there was a significant positive correlation between mean and difference MPOD values.Conclusions If MPOD is being monitored over time then any change less than 0.58 units should not be considered clinically significant as it is very likely to be due to instrument noise. The size of the coefficient appears to be positively correlated with MPOD.