TY - JOUR T1 - Assessment of a new Goldmann applanation tonometer JF - British Journal of Ophthalmology JO - Br J Ophthalmol SP - 42 LP - 46 DO - 10.1136/bjo.2010.182469 VL - 96 IS - 1 AU - M Egli AU - D Goldblum AU - A Kipfer AU - K Rohrer AU - C Tappeiner AU - M Abegg AU - L Berger AU - A Schoetzau AU - M E Iliev Y1 - 2012/01/01 UR - http://bjo.bmj.com/content/96/1/42.abstract N2 - Background/aims The classic Goldmann applanation tonometer (GAT) has been further developed by Haag-Streit International. The applanation principle has been retained, while the internal force transmission and the pressure gauging have been optimised, the display of results digitised. The authors compared the GAT standard with the new GAT digital.Methods Four fixed tonometer pairs were used. The protocol included: non-contact pachymetry, slit-lamp examination, three consecutive measurements with each tonometer with a 5 min interval in between, check for side effects in 15 min. Three groups (intraocular pressure (IOP) levels) were defined: (1) IOP≤16; (2) IOP>16 and <23; (3) IOP≥23 mm Hg.Results 125 Patients (250 eyes) were evaluated. IOP (mm Hg), GAT standard versus GAT digital, for the rights eyes was: Group 1: 12.94±0.55 versus 13.11±0.53, p=0.71. Group 2: 18.26±0.59 versus 18.03±0.52, p=0.53; Group 3: 30.28±0.48 versus 30.42±0.41, p=0.97; all right eyes: 17.48±7.48 versus 17.73±7.4, p=0.99. For the left eyes, there was no significant difference, either. The correlation was very good and was not influenced by the IOP level. The Pearson coefficient for the right eye was 0.985, and for the left eye 0.994. In the Bland–Altman analysis, although there were two single readings that differed by as much as 5 mm Hg, GAT digital measures showed almost no skew, and the mean difference was 0.03±1.23 mm Hg (n=250). A multiple regression analysis showed no influence of order of measurement, eyeside or pachymetry.Conclusions The new GAT digital is as reliable and safe as GAT standard. IOP values correlate well. It offers a digitised display and a wireless transfer of data. The display of values up to the first decimal digit is not necessarily associated with a more precise measurement, but may offer an additional comfort compared with the 2 mm Hg scale of the classic GAT. ER -