RT Journal Article SR Electronic T1 Novel device for expanding Schlemm's canal: a morphological study JF British Journal of Ophthalmology JO Br J Ophthalmol FD BMJ Publishing Group Ltd. SP 875 OP 877 DO 10.1136/bjophthalmol-2014-305540 VO 99 IS 7 A1 Grieshaber, Matthias C A1 Stegmann, Robert A1 Grieshaber, Hans R A1 Meyer, Peter YR 2015 UL http://bjo.bmj.com/content/99/7/875.abstract AB This study was performed to assess any changes in Schlemm's canal (SC) in human autopsy eyes possibly resulting from the injection of a viscoelastic substance and implantation of a new device (the Stegmann Canal Expander, SCE) into SC. After the surgical procedure, eyes were fixed, dissected into quartered segments and examined using light and scanning electron microscopy. Tissue sections displayed a marked dilation of SC and of the collector channels compared with untreated control sections. The SC walls were disrupted in some areas due to viscodilation, but not due to SCE implantation. In all eyes, the entire 9-mm length of the SCE was positioned inside the canal, keeping the canal wide open. The SCE diameter of 240 μm remained unchanged after implantation. Injection of a viscoelastic substance into SC leads to marked dilation of SC and collector channels. Implanting the SCE into SC allowed persistent expansion of SC and stretching of the trabecular meshwork. This may increase the permeability of the trabecular meshwork, reduce its resistance to aqueous humour and maintain circumferential flow within SC. Experimental and clinical studies should determine the impact of this new canal expander in terms of lowering intraocular pressure in glaucoma surgery.