TY - JOUR T1 - Repeatability and agreement of ocular biometry measurements: Aladdin versus Lenstar JF - British Journal of Ophthalmology JO - Br J Ophthalmol SP - 1223 LP - 1229 DO - 10.1136/bjophthalmol-2016-309365 VL - 101 IS - 9 AU - Colm McAlinden AU - Rongrong Gao AU - Ayong Yu AU - Xiaorui Wang AU - Jing Yang AU - Ye Yu AU - Hao Chen AU - Qinmei Wang AU - Jinhai Huang Y1 - 2017/09/01 UR - http://bjo.bmj.com/content/101/9/1223.abstract N2 - Objective To assess the repeatability and agreement between the Aladdin and Lenstar biometers.Methods Three consecutive measurements of the right eye of 102 subjects were acquired by the Aladdin (V.1.1.3) and the Lenstar (V.1.0.3), respectively, with a random order of which biometer to use first. Parameters compared included axial length (AL), anterior chamber depth (ACD), flat keratometry (Kf), steep keratometry (Ks), mean keratometry (Km), J0, J45 and white-to-white (WTW) distance. Intraocular lens (IOL) power formulas included SRK/T (Sanders-Retzlaff-Kraff/Theoretical), Holladay 1, Hoffer Q and Haigis. Repeatability (Sr), repeatability limit (r) and Bland-Altman limits of agreement (LoA) were calculated.Results Sr was marginally better with the Lenstar for AL (0.02 mm), ACD (0.02 mm) and WTW distance (0.07 mm), but Sr was marginally better with the Aladdin for Kf (0.10 dioptre (D)), J0 (0.07 D) and J45 (0.06 D). In terms of agreement, the mean difference was small and non-statistically significant between the two devices for ACD, Ks and J45. The mean difference for Kf, Km, J0 and WTW distance were all statistically significantly lower with the Aladdin than the Lenstar for the exception of WTW distance. The Bland-Altman LoA displayed narrow ranges implying good agreement with the exception for WTW distance. For IOL formulas, the mean difference was very small and not statistically significantly different. The LoA were narrow (within 0.50 D), implying acceptable use of each device interchangeably.Conclusions High levels of repeatability and agreement were found between the Aladdin and Lenstar suggesting that the devices may be used interchangeably in normal eyes with refractive error. ER -