Table 2 Validation study data using goggles 6 to 13
GoggleVertical width (mm)Horizontal width (mm)Face area (mm2)Rubber width (mm)Rubber depth (mm)mm Hg IOP rise (2 min)mm Hg IOP rise (20 min)
Mean (SD)n (95% CI)Mean (SD)n (95% CI)
6274211866.34.25.3 (1.9)4 (1.8)3.0 (3.0)4 (3.0)
732479857.16.35.0 (4.2)6 (3.3)5.5 (4.5)6 (3.6)
827467917.75.55.3 (4.3)6 (3.5)4.9 (2.7)6 (2.1)
923447967.111.714.7 (11.7)6 (9.3)13.4 (11.2)6 (9.0)
1028368918.75.56.5 (4.9)6 (4.3)7.2 (4.8)6 (3.9)
11345213899.05.33.6 (4.4)7 (3.3)3.4 (4.2)7 (3.1)
12356517719.49.51.5 (0.7)4 (0.7)1.5 (0.8)4 (0.8)
135362258113.618.50.4 (0.5)4 (0.5)0.4 (0.6)4 (0.6)
  • The goggle dimensions and mean intraocular pressure (IOP) rise induced in the right eyes of each of these goggles are presented. The mean IOP drop from 2 min to 20 min after goggle application was insignificant (pā€Š=ā€Š0.91).