TableĀ 2

The differences in measurements between IOLMaster 500 and IOLMaster 700; n (%)

Keratometry measurements (n=188)
Anterior chamber depth (n=175)Axial length (n=188)K1K2
Depth (mm)n (%)Length (mm)n (%)Keratometry measurements (D)n (%)Keratometry measurements (D)n (%)
<0.1 mm90 (52)No difference43 (24)<0.5D181 (96)<0.5D173 (92)
<0.2 mm154 (88)<0.02 mm102 (56)<1D188 (100)<1D188 (100)
<0.3 mm175 (100)<0.03 mm150 (82)
<0.05 mm165 (91)
<0.10 mm177 (96)
<0.15 mm188 (100)