Table 1

Deviation between EYEMATE-SC and Goldmann tonometry presented with corneal astigmatism values at the different time points

VisitMean of paired delta
(mm Hg)
Lower LoA
(mm Hg)
Upper LoA
(mm Hg)
Subj. CYL mean (dpt)Subj. CYL STD (dpt)Subj. CYL min (dpt)Subj. CYL max (dpt)
Screening−1.11.5−4.80.0
D11.5−3.36.3−1.31.3−5.0−0.3
D31.2−6.28.7−2.11.9−7.8−0.3
D102.5−5.510.5−2.31.6−6.5−0.3
D301.9−4.78.5−1.41.2−5.3−0.3
D900.1−5.55.6−1.41.2−5.50.0
  • CYL, cylinder; LoA, limit of agreement; STD, standard deviation.