Defects in disposable contact lenses compromise ocular integrity

https://doi.org/10.1016/0892-8967(92)90027-BGet rights and content

Abstract

To ascertain the prevalence of defects in disposable contact lenses, an Optimec JFC contact lens analyser (× 16 magnification) was used to examine 150 Acuvue, 150 NewVues, and 150 SeeQuence lenses, all −3.00D in power. The percentage of each lens type found to be defective was: Acuvue 75%, NewVues 5%, and SeeQuence 9%. A single-center, simultaneously-controlled, double-masked, randomized study was conducted to determine whether such defects compromise ocular integrity. Each of 29 subjects wore a defective Acuvue lens in one eye and a non-defective Acuvue (control) lens in the other eye on an extended-wear basis for one week. Lenses with defects visible to the naked eye were excluded. Although there was no significant difference in vision, subjective sensation, or bulbar conjunctival hyperemia between eyes wearing defective vs. control lenses, a significantly greater ocular response was observed in eyes wearing defective lenses in the form of corneal epithelial microcysts (p = 0.04), corneal staining (p = 0.02), and conjunctival staining (p = 0.04). Two cases of peripheral corneal infiltrates were observed; both cases were unilateral and occurred in eyes wearing defective lenses. Such compromise may predispose the eye to more serious complications such as infection and corneal ulceration.

References (33)

  • D.K. Devries et al.

    A clinical evaluation of edge induced conjunctival staining with Acuvue SeeQuence [sic] disposable lenses

    Optom Vis Sci

    (1989)
  • N.J. Lingel et al.

    A clinical evaluation of edge induced conjunctival staining with SeeQuence™ and NewVues™ disposable lenses

    Optom Vis Sci

    (1990)
  • B.A. Weissman et al.

    Oxygen transmissibility of disposable hydrogel contact lenses

    Am J Ophthalmol

    (1991)
  • M.J.A. Port

    The Optimec Contact Lens Analyser

    Optician

    (1981)
  • G.E. Lowther

    Hydrophilic lens inspection with phase contrast microscopy

    Am J Optom Physiol Opt

    (1981)
  • C.W. McMonnies et al.

    Assessment of conjunctival hyperemia in contact lens wearers. Part 1

    Am J Optom Physiol Opt

    (1987)
  • Cited by (47)

    • Complications

      2023, Contact Lens Practice, Fourth Edition
    • Reusable Lenses

      2023, Contact Lens Practice, Fourth Edition
    • Soft Lens Measurement

      2023, Contact Lens Practice, Fourth Edition
    • Soft Lens Manufacture

      2023, Contact Lens Practice, Fourth Edition
    • Statistical Methods

      2021, Statistical Methods
    • Soft Lens Measurement

      2018, Contact Lens Practice
    View all citing articles on Scopus
    View full text