Comparison of anterior segment optical tomography parameters measured using a semi-automatic software to standard clinical instruments

PLoS One. 2013 Jun 4;8(6):e65559. doi: 10.1371/journal.pone.0065559. Print 2013.

Abstract

Objective: To compare anterior segment parameters measured using a semi-automatic software (Zhongshan Angle Assessment Program, ZAP) applied to anterior segment optical coherence tomography (AS-OCT) images, with commonly used instruments.

Methods: Cross-sectional study of a total of 1069 subjects (1069 eyes) from three population-based studies of adults aged 40-80 years. All subjects underwent AS-OCT imaging and ZAP software was applied to determine anterior chamber depth (ACD), central corneal thickness (CCT), anterior and keratometry (K) - readings. These were compared to auto-refraction, keratometry and ocular biometry measured using an IOLMaster, ultrasound pachymeter and auto-refractor respectively. Agreements between AS-OCT (ZAP) and clinical instrument modalities were described using Bland-Altman, 95% limits of agreement (LOA).

Results: The mean age of our subjects was 56.9±9.5 years and 50.9% were male. The mean AS-OCT (ZAP) parameters of our study cohort were: ACD 3.29±0.35 mm, CCT 560.75±35.07 µm; K-reading 46.79±2.72 D. There was good agreement between the measurements from ZAP analysis and each instrument and no violations in the assumptions of the LOA; albeit with a systematic bias for each comparison: AS-OCT consistently measured a deeper ACD compared to IOLMaster (95% LOA -0.24, 0.55); and a thicker CCT for the AS-OCT compared to ultrasound pachymetry (16.8±0.53 µm 95% LOA -17.3, 50.8). AS-OCT had good agreement with auto-refractor with at least 95% of the measurements within the prediction interval (P value <0.001).

Conclusion: This study demonstrates that there is good agreement between the measurements from the AS-OCT (ZAP) and conventional tools. However, small systematic biases remain that suggest that these measurement tools may not be interchanged.

Publication types

  • Comparative Study
  • Randomized Controlled Trial

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Automation
  • Female
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Male
  • Middle Aged
  • Reference Standards
  • Software*
  • Tomography, Optical / instrumentation
  • Tomography, Optical / methods*