Comparison of the accuracy of the Binkhorst, Colenbrander, and SRK implant power prediction formulas

J Am Intraocul Implant Soc. 1981 Fall;7(4):337-40. doi: 10.1016/s0146-2776(81)80031-6.

Abstract

The prediction accuracy of the Binkhorst, Colenbrander, and SRK formulas were compared in five series from different sources totalling 654 cases. The SRK formula was superior to both the other formulas by having a smaller average error per case, a smaller range of error from highest minus to highest plus, and a smaller proportion of cases with greater than 2 diopters (D) of error in all five series studied. In four of the five series, the SRK formula also had the greatest proportion of cases with less than 1 D of error. The Colenbrander formula was superior to the Binkhorst formula in all five series with regard to average error, proportion of cases with less than 1 D of error, and proportion of cases with more than 2 D of error. The range of error from highest minus to highest plus was equivalent with the two formulas. All available published literature on the accuracy of implant power prediction formulas was reviewed and appears to support our findings.

MeSH terms

  • Humans
  • Lenses, Intraocular / standards*
  • Visual Acuity*